Beer-Lambert law (source code)

= Beer-Lambert law
{c}
{title2=$A=\varepsilon lc$}
{wiki=Beer–Lambert_law}

The <Beer-Lambert law> relates <absorbance> $A$ to an absorbing species' <concentration> $c$ and path length $l$ by $A=\varepsilon lc$ in a homogeneous medium using sufficiently narrow-band light. The coefficient $\varepsilon$ depends on the species, <wavelength> and <chemical solution> conditions. This makes <Ellman's reagent> useful for quantitative <thiol> assays once its released colored species has been calibrated. <IUPAC Compendium of Chemical Terminology, fifth edition (Gold Book) by International Union of Pure and Applied Chemistry/Beer–Lambert law>[IUPAC Compendium of Chemical Terminology, fifth edition (Gold Book) by International Union of Pure and Applied Chemistry; IUPAC Gold Book: Beer-Lambert law and validity conditions]{ref}

Here $c$ is <molar concentration>, $l$ is the geometric <length> through the sample, and $\varepsilon$ is the <molar absorption coefficient>. The ideal beam has parallel <light rays>, a single <wavelength>, and a medium with uniform composition and <isotropy>; finite spectral width must be small relative to features in the <optical spectrum>.<IUPAC Compendium of Chemical Terminology, fifth edition (Gold Book) by International Union of Pure and Applied Chemistry/Beer–Lambert law>[IUPAC Compendium of Chemical Terminology, fifth edition (Gold Book) by International Union of Pure and Applied Chemistry; Beer–Lambert law, entry B00626: opening definition, equations and spectral-bandwidth qualification]{ref}

For a <spectrophotometer>, distinguish the source's output from the reference that reaches the <photodetector>. Let $P_0$ be the received <radiant power> through a <reference blank> and $P$ the received <radiant power> through the sample, after subtracting the <photodetector>'s dark response. Both refer to the same selected <wavelength> band and collection geometry. Define the blank-normalized <transmittance> and <absorbance> by
$$
T=\frac{P}{P_0},\qquad A=-\log_{10}T,\qquad T=10^{-A}.
$$
The <reference blank> includes the <solvent> and other assay ingredients except the absorbing target. Matching <cuvettes> and conditions makes their common losses divide out. Thus this $T$ is relative to the <reference blank>; it is not the absolute <transmittance> of the entire <cuvette> and its contents. Both <radiant powers> can be expressed in <watts>, while their <ratio>, $T$, and its <logarithm>, $A$, are <dimensionless>. An instrument displaying $40\%$ transmission means $T=0.40$, not $T=40$.

\Image[/beer-lambert-reference-sample-detector.png]
{title=The <reference blank> and sample establish a <radiant power> <ratio>}
{id=image-beer-lambert-reference-sample-detector}
{description=Original schematic of two successive measurements in the same <spectrophotometer>, not a drawing of a particular commercial instrument. The <monochromator> selects the same <wavelength> band; the same <photodetector> reads the <reference blank> and sample. The marked $l$ is geometric <length> through the <chemical solution>, distinct from <optical path length>. Arrow widths are schematic, not measured <radiant powers>.}

Why take a <logarithm>? Consider two consecutive identical absorbing layers. Each transmits a fraction $t$ of the <radiant power> entering it, so together they transmit $t^2$. Their <absorbances> add because $-\log_{10}(t^2)=2(-\log_{10}t)$. Doubling the sample <length> doubles <absorbance> while squaring <transmittance>. This turns multiplicative loss into an additive scale and makes a <straight line> useful for inferring <molar concentration>.

A short model derivation makes the assumption explicit. Suppose a thin layer of thickness $dx$ removes the fraction $kc\,dx$ of its incoming <radiant power>, where $k$ is a fixed <constant of proportionality> and $c$ is a uniform <molar concentration>. The corresponding <ordinary differential equation>, initial condition, and solution are
$$
\frac{dP}{dx}=-kcP,\qquad P(0)=P_0,\qquad
P(l)=P_0 e^{-kcl}.
$$
Taking the base-ten <logarithm> gives $A=kcl/\ln 10$, so $\varepsilon=k/\ln 10$. The <exponential decay> follows from equal layers removing equal fractions of what remains. Proportionality to <molar concentration> additionally assumes that the absorbing <chemical species> retains the same response as its <molar concentration> changes; the geometric argument alone does not establish that.