Detection-threshold selection bias in pulse-energy means occurs when detected pulses have a different energy distribution from all emitted pulses. A detector conditions the reported mean on detection, not merely on emission. Let be emitted-pulse energy with finite mean and let be detection probability conditional on emission . ThenThe denominator must be positive. An ideal energy threshold has above threshold and zero below. Its nondecreasing selection preferentially retains stronger pulses; false triggers or selection on another variable require a different model.
For an original distribution, emitted energies are , and with probabilities , and . Their conditional pulse-energy mean given emission is . A perfect receiver accepting only energies above detects of emitted pulses and reportsThe detected-event mean is about too high for the emitted-pulse mean, even with perfect energy measurements on every accepted event.
A trigger counter, a lower-threshold reference monitor and a threshold sweep can distinguish missing detections from missing emissions. Historical calorimetric inference uses total energy, exposure time and pulse rate; the rate must describe the population whose mean is requested.[ref]
Calibration Service for Laser Power and Energy at 248 nm by Rodney W. Leonhardt. §6.2, Laser Stability, calorimetric pulse-energy inference from energy, injection time and pulse rate
This selection example is an independent probability model, not a reconstruction of that experiment.Known positive detection probabilities can support inverse probability weighting. A hard threshold with zero probability below it leaves that part of the energy distribution unidentified from detected events alone. Reporting the acceptance fraction as well as the detected-event mean makes the conditioning explicit; it does not recover the unseen weak-pulse energies.
Missing-pulse bias in average-power inference arises when a nominal trigger rate is used as though every trigger emitted a pulse. With background removed and complete pulses counted, let be the number of triggers and the number of emissions during . ThenThe quotient is mean energy per trigger, including failed triggers as zero-energy outcomes. It equals the conditional pulse-energy mean given emission only when every trigger emits, or after correction for the measured emission fraction. Historical calorimetric inference uses energy, injection time and pulse rate; those quantities must describe the same collection.[ref]
For an original five-trigger record, suppose energies are . Mean energy per trigger is , while the two emitted pulses average . At a nominal trigger rate, their average power is . The actual emission rate is , consistent with the same power and emitted-pulse mean.
A pulse monitor with a threshold above the weaker pulse could detect only the event. Its detected-event average would then be , also different from the true emitted-pulse mean. Failure to detect is not automatically failure to emit. A calibrated pulse-resolving receiver and a trigger counter distinguish emitted, detected and commanded events. The average power of a laser pulse train article additionally treats continuous backgrounds and observation-window accounting.