= Electropherogram

An \b[<electropherogram>] is a recorded trace of an <electrophoresis> separation. In <capillary electrophoresis>, it commonly plots detector <signal> against migration time: separated populations reach a detection window at different times and produce peaks. Peak position, width, height and integrated area describe different features of that trace.

The position depends on the migration mechanism and geometry. In a simple uniform-field model with detector distance $L_d$, <electric field> $E$, <electrophoretic mobility> $\mu_{\rm ep}$ and <electroosmotic flow> represented by mobility $\mu_{\rm eo}$,
$$
v=(\mu_{\rm ep}+\mu_{\rm eo})E,\qquad t_m=\frac{L_d}{v},
$$
when the resulting drift is towards the detector. A changed buffer-flow contribution can therefore shift migration times without changing the analyte's identity. The instrument's actual timing and mobility conventions must be specified; the first-party software documentation gives its explicit migration-time and apparent-mobility definitions.<Processing Methods - CE data by Waters>[Procedure, Step 1 and Step 5, items 1 and 6]{ref}

An <electropherogram peak area> is the integral above an assigned baseline. For an original Gaussian-component model,
$$
y(t)=b+\sum_i\frac{A_i}{\sigma_i\sqrt{2\pi}}
\exp\left[-\frac{(t-t_i)^2}{2\sigma_i^2}\right].
$$
The component area is $A_i$, while its height above baseline is $A_i/(\sigma_i\sqrt{2\pi})$. Choose $(A_1,\sigma_1,t_1)=(1,0.4,5)$ and $(A_2,\sigma_2,t_2)=(2,0.8,10)$ in compatible signal/minute units. The two component heights are equal, although the second has twice the integrated signal. These are synthetic components, not experimental analyte concentrations.

\Image[/electropherogram-peak-area.png]
{title=An original synthetic electropherogram separates peak height, integrated signal and migration time}
{description=The baseline-subtracted Gaussian areas are 1 and 2 signal-minutes, with widths 0.4 and 0.8 minutes and centers at 5 and 10 minutes. The component heights are equal. No chemical identity or absolute amount is inferred from the modeled trace.}

The cited software defines time-corrected area as $A_i/t_i$.<Processing Methods - CE data by Waters>[Procedure, Step 5, items 4–5]{ref} In this example, both corrected areas are $1/5=2/10=0.20$. This is another observable, not permission to equate every corrected area with an absolute amount without response calibration and an appropriate measurement model.

A peak can also contain unresolved co-migrating species. Its apparent simplicity does not establish chemical purity. Useful interpretation therefore combines the trace with baseline selection, migration references, detector response and independent identification evidence, rather than treating each visible peak as a complete chemical explanation.