= Enzyme assay
An <enzyme assay> measures the catalytic activity of an <enzyme> by following consumption of a <substrate> or appearance of a reaction product under specified conditions. The signal may be chemical concentration directly or a calibrated optical readout, such as <absorbance> or <fluorescence>.
For example, if product concentration increases from 0 to 6 micromolar in the first two minutes, a linear initial interval gives a rate of 3 micromolar per minute. A no-enzyme control tests whether the signal also changes without the catalyst. Comparing rates requires the same <temperature>, <pH>, <substrate> concentration and detection conditions; raw brightness alone is not an activity unit.
The initial interval matters because later measurements can be affected by substrate depletion, product accumulation or loss of enzyme activity. Assay-development guidance therefore recommends establishing the linear initial-rate region before comparing inhibitors or <enzyme> preparations.<Basics of Enzymatic Assays for HTS by Harold B. Brooks and others>[Initial velocity, linear reaction range, substrates and controls]{ref}