Everhart–Thornley detector (source code)

= Everhart–Thornley detector
{c}

An <Everhart–Thornley detector> converts emitted <electrons> into an amplified optical signal in a <scanning electron microscope>. A positively biased collector attracts low-energy <secondary electrons> toward a <scintillator>; their impact produces <light> carried through a light guide to a <photomultiplier>.

The measured brightness depends on how many emitted <electrons> reach the collector, not just on how many leave the surface. Changing specimen tilt or collector bias can therefore change an image without changing composition. Comparing such settings helps identify collection-geometry effects in <secondary-electron topographic contrast>. https://www.jeol.com/words/semterms/20121024.070858.php[JEOL manufacturer glossary, SED: scintillator, light guide and photomultiplier arrangement]{ref}