= Extracted ion profile
An \b[extracted ion profile] plots selected <mass-to-charge ratio> signal across a time-ordered series of <mass spectra>. The selection can be one value or a set of values. When time follows a chromatographic separation, the result is an <extracted ion chromatogram>; the broader profile term also covers other time-dependent sampling.<Extracted ion profile in IUPAC Gold Book>[Entry12410, definition and sampling examples.]{ref}
For recorded spectra $I_k(t_j)$, an explicit <extracted ion mass window> $W$ gives the reconstructed trace
$$
E_W(t_j)=\sum_{k\in W}I_k(t_j).
$$
This selects information already present in the spectra. It does not create missing mass measurements or turn the acquisition into a different instrumental measurement. Unlike a <total ion current profile>, it excludes signal outside the stated window.
An original three-scan example shows how the window changes the result. At mass coordinates $499.98$, $500.00$ and $500.03$, let the intensity vectors be $(0,1,0)$, $(3,8,5)$ and $(0,2,10)$ at successive times. A window $500.00\pm0.01$ selects only the middle coordinate and gives $(1,8,2)$. Widening to $500.00\pm0.04$ includes all three coordinates and gives $(1,16,12)$. The larger late signal in the broad trace comes from added spectral content; it is not evidence that the middle-coordinate signal remained high.
Choosing a window requires considering <mass resolving power>, mass-coordinate accuracy and what the instrument actually recorded. A narrow window can exclude a shifted target peak; a wide window can include neighbors. Even an isolated mass coordinate need not identify a unique chemical species, because different ions can share it. Preserve the raw spectra and state the window, summation rule and time axis so the profile can be reproduced and interpreted.
Back to article page