Fixed-cistern barometer (source code)

= Fixed-cistern barometer

A <fixed-cistern barometer> is a <mercury barometer> whose reservoir geometry is left fixed during a reading. As mercury rises in the narrow tube, it is withdrawn from the reservoir and the reservoir level falls. The measured column-height change consequently exceeds the upper-surface displacement relative to the instrument frame.<Mercury Barometers and Manometers by W. G. Brombacher and others>[§5.3, fixed cistern barometers and manometers]{ref}

For constant tube and reservoir <areas> $A_t$ and $A_c$, conservation of mercury <volume> gives a reservoir drop $(A_t/A_c)\Delta z$ when the tube surface rises by $\Delta z$. Thus $\Delta h=(1+A_t/A_c)\Delta z$ in this ideal geometry. A calibrated scale can incorporate that ratio; temperature changes require additional corrections. A <Fortin barometer> instead restores the lower surface to a fixed index before reading, illustrating two ways of controlling the same reference-level problem.