= FRET optical crosstalk
{c}
<FRET optical crosstalk> includes <FRET donor> emission leaking into the <FRET acceptor> detection band and direct excitation of the <FRET acceptor> by the nominal <FRET donor> excitation light. Measure donor-only, acceptor-only and unlabeled preparations with the same <fluorescence microscope> settings to separate these contributions and background.<Advanced Fluorescence Microscopy Techniques—FRAP, FLIP, FLAP, FRET and FLIM by Hellen C. Ishikawa-Ankerhold and others>[§3.4.2, samples i–iii and equations 20–21: spectral correction controls]{ref}
An explicit linear measurement model makes the controls' purpose clear. Let $D$ be background-subtracted donor-band counts under donor excitation, $A$ acceptor-band counts under acceptor excitation, and $F$ acceptor-band counts under donor excitation. Donor-only and acceptor-only preparations determine leakage coefficients $a$ and $b$. Then
$$
S=F-aD-bA
$$
is a corrected <FRET sensitized emission> signal under the assumed linear, fixed optical response. In a synthetic example $D=100$, $A=80$, $a=0.10$, $b=0.05$ and $F=34$ give $S=20$. Even a nontransferring sample would give $F=14$ from these two contaminants. Correcting that signal still does not convert $S$ into <FRET efficiency> without a calibration of excitation, yields and detection.
A biological comparison also needs a pairing-negative preparation and a known transferring positive preparation. Match label abundance and localization where possible; changing <concentration> can change encounter frequency as well as brightness. Check that attaching the labels preserves the <protein> function being investigated. Otherwise the measurement can faithfully report a perturbation introduced by the labels themselves.
Back to article page