George C. Rhoderick Created 2026-10-14 Updated 2026-11-05
George C. Rhoderick coauthored research validating microwave spectroscopy for trace-gas analysis. The work tests quantitative response against prepared gas mixtures and calibration standards.
Such comparisons connect a measured spectral amplitude with a chemical amount. Sample delivery and expansion can influence that connection even when line assignment is secure.[ref]
The Validation of Fourier-Transform Microwave Spectroscopy for Trace-Gas Analysis. Printed pages 25–27, quantitative validation
Identity source: www.nist.gov/publications/search_by_author/1148686
By Pamela M. Chu, R. D. Suenram, R. S. DaBell, Angela R. Hight Walker, George C. Rhoderick, Gerald T. Fraser.
Online text: nvlpubs.nist.gov/nistpubs/Legacy/IR/nistir6816.pdf
Authored extended-abstract contribution, printed pp. 25–28, with the six-person byline on p. 25. The institutional report was issued by NIST in February 2002; the issue date is distinct from the original workshop date. The cover credit to J. C. Boudreaux does not establish authorship of this contribution or editorship of the report.
Cited passages: The Validation of Fourier-Transform Microwave Spectroscopy for Trace-Gas Analysis, printed pp. 25–26, Figures 2–3: coaxial apparatus, excitation and OCS readout; PDF pages 31–32; printed pp. 26–27: nozzle/carrier comparison in Figure 6, blind mixture in Table II, one-standard calibration and nozzle-expansion limitations.