= Industrial platinum resistance thermometer

= IPRT
{c}
{synonym}

An <industrial platinum resistance thermometer> is a <platinum resistance thermometer> built for practical measurement conditions. The label <Pt100> specifies a nominal reference <electrical resistance>, not an <SPRT>, a calibration certificate, or a universal accuracy. Calibrate the probe together with its readout when the desired result is the performance of the complete indicating system.