= Polarimetry
<Polarimetry> measures <optical rotation> with a <polarimeter>. The vertical Vernier Go Direct instrument uses a $589\,\mathrm{nm}$ <light-emitting diode> source, fixed <polarizer>, sample cell, motor-driven <polarimeter analyzer>, <photodetector> and optical <rotary encoder>. Calibrate using the same <solvent> at $10\,\mathrm{cm}$ fill height; rinse with sample, fill to $10.0\,\mathrm{cm}$, and read the meniscus to $0.1\,\mathrm{cm}$. The readout plots illumination against analyzer <angle>; a maximum near zero or a cosine-squared fit locates $\alpha$.<Go Direct Polarimeter User Manual by Vernier>[Using the Product steps 2–4; Specifications; How the Sensor Works]{ref}
An actual teaching experiment in the Truman State University-hosted Optical Activity Laboratory handout, revised July 2013, uses <sucrose> <chemical solutions>. Weigh the <sucrose>, dissolve completely without heating, and measure final solution <volume (thermodynamics)> for <concentration>. Use one <polarimeter> throughout; measure the <water> blank's extinction reading, then each sample's extinction reading and actual liquid path length near $10\,\mathrm{cm}$. Subtract the blank and plot corrected <optical rotation> against <concentration>. The attached apparatus description specifies a vertical <light> path, an incandescent or tungsten source with yellow filter, sheet <polarizers> and a rotating eyepiece <polarimeter analyzer>. Its coarse demonstration permits <temperature> control to the nearest $5^\circ\mathrm C$, path length to the nearest millimeter and approximately $1^\circ$ angular precision. These are design tolerances, not measured results.<Optical Activity Laboratory, Truman State University>[Optical Activity Laboratory, revised July 2013, pp. 1–2 Experiment 1; p. 3 The Design and General Use of Simple Polarimeters]{ref}
For a quantitative comparison, equilibrate blank, reference and sample at the same recorded <temperature>; use a thermostatted cell when needed. This is a recommended control, not a claim that the teaching instruments provide a thermostat. A yellow filter also passes a band of <wavelengths>, so its result need not match a narrow-source reference exactly. The laboratory's concentration series tests the linear model rather than establishing it in advance; a second series varying only path length tests the other factor.
For our original ideal model, <Malus's law> gives
$$
I(\theta)=I_0\cos^2(\theta-\alpha),
$$
where $I_0$ is <light intensity> immediately before the ideal <polarimeter analyzer>, after sample transmission. It is not the intensity of the original unpolarized source. The <angle> difference is evaluated in consistent units; the plotted degree values are converted to <radians> before computing the <cosine>. Assume transmitted <light> retains <linear polarization>, with negligible scattering and no polarization-dependent absorption. The pure positive model, $75{:}25$ mixture and <racemate> shift the maxima to $8^\circ$, $4^\circ$ and $0^\circ$, respectively. Minima shift by the same amount, to $98^\circ$, $94^\circ$ and $90^\circ$. Thus an extinction measurement and a maximum measurement recover the same shift after their respective blanks are subtracted.
\Image[/polarimetry-mixtures-and-analyzer-intensity.png]
{title=Computed <optical rotations> and <polarimeter analyzer> intensity for a stipulated <enantiomer> pair}
{id=image-polarimetry-mixtures-and-analyzer-intensity}
{description=Original model, not experimental data: $k=40\,\mathrm{deg\,mL\,g^{-1}\,dm^{-1}}$, $l=1\,\mathrm{dm}$ and total pair <concentration> $c=0.200\,\mathrm{g\,mL^{-1}}$. Left: additive <optical rotation> versus positive-member <mole fraction>. Right: normalized <Malus's law> curves for $\alpha=8^\circ,4^\circ,0^\circ$, with a close view of their extinction minima. Normalization by each trace's pre-analyzer $I_0$ compares orientations rather than sample transmission losses.}
Repeat blanks and use a reference to check drift and scale. Bubbles, particles, stressed windows and sample absorption can spoil the ideal curve. Vernier specifies $\pm1^\circ$ accuracy; the manual's $\pm0.1^\circ$ fit reproducibility does not establish that accuracy. It warns against interpreting its illumination readout quantitatively and requires a transparent homogeneous sample.<Go Direct Polarimeter User Manual by Vernier>[Using the Product, cosine-squared fit; Specifications; Troubleshooting, illumination and sample limits]{ref} Neither a zero reading nor a matching <specific rotation> uniquely identifies a <molecule>.