= Scanning tunneling microscope
{wiki}
= Scanning tunnelling microscope
{synonym}
= STM
{c}
{synonym}
In the 1982 experiment, separate drives using <piezoelectricity> scanned a <metal> tip laterally and adjusted its height. A control unit held <electric current> fixed at constant tip-sample <electric potential difference>; calibrated drive voltages supplied the displacement readout. Their Fig. 1 explicitly contrasts a surface step with apparent height from contamination changing the <work function>.<Surface Studies by Scanning Tunneling Microscopy by Gerd Binnig and others>[pp. 58–59, Fig. 1, equation (1), piezodrive calibration]{ref}
<Gerd Binnig> and <Heinrich Rohrer> describe a later <vibration isolation> system using two spring stages and <eddy current> <damping> inside a <vacuum> chamber, with <sputtering> and <annealing> for surface preparation. These measures addressed unwanted motion and surface condition separately.<Scanning Tunneling Microscopy: From Birth to Adolescence by Gerd Binnig and others>[p. 398, spring suspension, eddy-current damping and sample preparation]{ref}
\Image[/scanning-tunneling-microscope-feedback.png]
{title=How a scanning tunneling microscope turns current feedback into an image}
{description=Original functional schematic of <constant-current scanning tunneling microscopy>. The loop retracts the tip when measured <electric current> exceeds its target. Recorded $z(x,y)$ includes <electronic contrast in scanning tunneling microscopy>; the diagram is not to scale.}