Time-of-flight atom detection (source code)

= Time-of-flight atom detection

Time-of-flight atom detection measures particles after releasing a trapped atomic cloud and allowing it to expand for a known time. For ballistic motion, $\mathbf r(t)=\mathbf r_0+\mathbf p_0t/m+\mathbf g t^2/2$, so after accounting for gravity the late position largely reflects the initial <momentum> when the initial cloud size is sufficiently small.

A position- and time-resolving detector can record individual arrivals and form spatial pair correlations. In <fermionic atom antibunching>, such correlations probe suppression of nearby detection pairs for the prepared state. Finite source size, interactions during early expansion, detector resolution and detection efficiency limit a literal initial-momentum interpretation; expansion is part of the experimental model, not merely a delay before measurement. A <microchannel plate detector> is one implementation for suitable metastable atoms.