= Vascular graft
A \b[vascular graft] is a conduit used to replace or reconstruct a <blood vessel> or to create a new flow route. A <bypass graft> is one application: the conduit directs <blood flow> around a diseased segment. The graft can use biological tissue or a manufactured material; a <tissue engineering> approach attempts to build a vessel substitute with living cells and supporting matrix.
The engineering requirements are simultaneous. The lumen must permit flow, the wall must withstand loading, the connections must remain intact, and the blood-contacting surface must interact appropriately with circulating blood. <Graft patency> records whether the route remains open; it is a different endpoint from the wall's strength or its appearance under <microscopy>.
A simplified thin-walled circular tube illustrates one mechanical demand. Balancing the pressure force on a lengthwise half-cylinder against circumferential wall tension gives <hoop stress>
$$
\sigma_\theta\simeq\frac{\Delta p\,r}{h},
$$
where $r$ is lumen radius, $h$ wall thickness and $\Delta p$ the pressure difference across the wall. For stipulated $r=2\,\mathrm{mm}$, $h=0.20\,\mathrm{mm}$ and $\Delta p=13.3\,\mathrm{kPa}$, this gives approximately $133\,\mathrm{kPa}$. The approximation assumes a thin wall and simple geometry; it does not predict the strength of a suture connection or a real multilayer graft.
A primary engineered-vessel study by <Laura E. Niklason> and colleagues measured static and dynamic compliance on a bench using controlled pressures and video-based tracking of vessel diameter. Separate tests assessed rupture strength. These apparatus measurements distinguished resistance to expansion, response during pulsatile loading and failure pressure.<Morphologic and mechanical characteristics of engineered bovine arteries by Laura E. Niklason and others>[Compliance measurements, printed pp.631–632; Figure9, p.636]{ref}
That distinction motivates <vascular graft compliance mismatch> and <vascular graft burst pressure> as separate assessments. A conduit can be strong at rupture yet differ mechanically from the vessel it joins. Conversely, surviving an initial pressure test does not establish long-term <graft patency> or biological remodeling. The measurement, tested material, loading history and follow-up interval must accompany claims about graft function.