Balanced homodyne detection

ID: balanced-homodyne-detection

Mix a signal mode and an independent coherent reference mode on a balanced beam splitter, detect both outputs, and subtract the currents. For and , the difference is . A reference gives mean with . Other beam-splitter conventions shift this phase calibration. The exact finite-reference balanced homodyne variance quantifies the strong-reference approximation.
Matched subtraction cancels the shared total-power term. It therefore suppresses common reference intensity noise, while imbalance and phase fluctuations reduce this advantage. Repeated samples measure a distribution, not just its mean; scanning reference phase supports optical homodyne tomography.