Manufacturer technical manual, Chapter 7, Technical overview: printed p. 102, About fluidics, describes laminar sheath flow and hydrodynamic focusing; p. 103 introduces optics. Printed p. 115, pulse processing: height is maximum digitized intensity; area integrates digitized samples over the acquisition window; width is separately measured. These definitions are instrument-specific implementations of pulse properties.