Detection-threshold selection bias in pulse-energy means

ID: detection-threshold-selection-bias-in-pulse-energy-means

Detection-threshold selection bias in pulse-energy means occurs when detected pulses have a different energy distribution from all emitted pulses. A detector conditions the reported mean on detection, not merely on emission. Let be emitted-pulse energy with finite mean and let be detection probability conditional on emission . Then
The denominator must be positive. An ideal energy threshold has above threshold and zero below. Its nondecreasing selection preferentially retains stronger pulses; false triggers or selection on another variable require a different model.
For an original distribution, emitted energies are , and with probabilities , and . Their conditional pulse-energy mean given emission is . A perfect receiver accepting only energies above detects of emitted pulses and reports
The detected-event mean is about too high for the emitted-pulse mean, even with perfect energy measurements on every accepted event.
A trigger counter, a lower-threshold reference monitor and a threshold sweep can distinguish missing detections from missing emissions. Historical calorimetric inference uses total energy, exposure time and pulse rate; the rate must describe the population whose mean is requested.[ref] This selection example is an independent probability model, not a reconstruction of that experiment.
Known positive detection probabilities can support inverse probability weighting. A hard threshold with zero probability below it leaves that part of the energy distribution unidentified from detected events alone. Reporting the acceptance fraction as well as the detected-event mean makes the conditioning explicit; it does not recover the unseen weak-pulse energies.