Solution

ID: past-exam-of-the-chemistry-course-of-the-year-4-of-the-natural-sciences-course-of-the-university-of-cambridge/2022/paper-3/section-b/38/b/solution

The hemispherical component needs 100.0 eV electrons, whereas electrons arrive at its entrance optics with 111.4 eV. The optics must therefore remove 11.4 eV, requiring a retarding-potential magnitude of 11.4 V. For a negative electron, moving to a more negative electrostatic potential increases potential energy, so the retarding region is 11.4 V negative relative to the analyser entrance. Equivalently the hemisphere entrance is 11.8 V negative relative to the sample vacuum reference after including the 0.4 V contact step. Reporting 11.8 V as the additional instrument retardation would count the contact correction twice.