A heated cathode emits electrons, accelerated through a potential difference onto a metal target. Deceleration produces a bremsstrahlung continuum with ; inner-shell ionization followed by electronic relaxation produces characteristic lines. The pictured instrument is consistent with the copper-target TEL-X-Ometer used for teaching diffraction. Its copper radiation has wavelength about , comparable with atomic plane spacings. A nickel filter preferentially suppresses copper , while slits collimate the beam; filtering does not remove every continuum photon.
A Geiger-Müller counter converts photon-induced gas ionization into countable electrical pulses. A scintillator with photomultiplier is another detector option. A detector-arm scan records count rate against scattering angle; coupled crystal rotation preserves the usual reflection geometry. These instrument features are documented by its manufacturer.[ref]
TEL-X-Ometer X-Ray System Manual by Joseph Dohm. Sections 3.5–3.6 and 7.1: crystal/detector geometry and detection; section 4.6: copper characteristic wavelengths
For the numerical calculations below use , explicitly identifying the characteristic line assumed. The photograph alone does not give a measured wavelength or angular travel limit; conclusions depending on these parameters are stated conditionally. A different selected line changes the angles but not a crystallographic extinction.