Equipartition trap calibration (source code)

= Equipartition trap calibration

A calibrated coordinate in a harmonic trap has equilibrium <variance> $\langle\delta x^2\rangle=k_BT/k$. This determines stiffness only after removing detector noise and accounting for drift, exposure averaging and finite bandwidth. Position <variance> does not supply a <force> calibration if the coordinate scale is unknown.