For Ørsted's experiment, place a long straight conducting wire close to and parallel with a freely moving horizontal compass needle. With no electric current, the needle follows the horizontal terrestrial magnetic field. Closing the circuit deflects it; reversing the electric current reverses the deflection, and moving the wire from above to below the needle also reverses the effect. For a long straight wire, the modern idealization is
with the circular direction given by the right-hand rule. These controls distinguish a magnetic action of the current from a fixed magnetized piece of wire or a mechanical disturbance.
Nobili's astatic galvanometer increases sensitivity in two ways. A coil with many turns multiplies the magnetic action of a given electric current. Two nearly equal magnetized needles are rigidly connected with opposite magnetic dipole moments and suspended by a fine thread. Their torques in the approximately uniform terrestrial field nearly cancel. In a common arrangement one needle lies inside the coil and the other above it, outside its strongest field, so the coil retains a substantial torque while the terrestrial restoring torque is reduced. A weak residual magnetic or suspension torque still provides a reproducible equilibrium; “astatic” means nearly cancelled terrestrial torque, not magnetic shielding. A museum instrument dated 1840 has many turns of fine wire and is identified with measurements of muscle currents.[ref]
For weak physiological sources, use a long fine-wire winding with many turns, rather than the short thick winding appropriate to a low-resistance thermoelectric source. The physiological source commonly has a large effective internal resistance; a moderate added coil resistance is then outweighed by the gain in turns. This is a source-and-detector matching argument, not a claim that arbitrarily much wire always improves sensitivity.
For example, at fixed mean turn length and fixed copper winding volume, . For source electromotive force and resistance , the small-angle driving torque is proportional to
Differentiation gives its maximum at . This simple optimal turn count for fixed winding volume model explains why high-source-resistance work favors more turns, while also exposing the limit. Actual instruments additionally trade response time, damping, geometry and source loading against their steady deflection.