A-mode ultrasound scan Created 2026-10-20 Updated 2026-11-05
An A-mode ultrasound scan plots the returning echo amplitude along one pulse-echo ultrasound beam. With assumed sound speed , echo delay is converted to depth by ; the factor two accounts for outward and return travel. Peaks mark reflecting interfaces along that line. [ref]
The horizontal coordinate is depth inferred from travel time, while the vertical coordinate is signal amplitude. An A-mode ultrasound scan therefore provides a one-dimensional depth profile rather than the spatial cross-section assembled by a B-mode ultrasound scan.
B-mode ultrasound scan Created 2026-10-20 Updated 2026-11-05
A B-mode ultrasound scan maps echo strength to displayed brightness and assembles multiple pulse-echo ultrasound lines into a spatial image. Each echo's location is inferred from beam direction and return time using an assumed sound speed. [ref]
A bright point represents a strong received echo, not a direct measurement of tissue density. Acoustic impedance contrasts, propagation losses and the instrument's display mapping all affect brightness. A moving image consists of repeated spatial frames; in an M-mode ultrasound scan, the second display axis instead tracks time along one beam.
M-mode ultrasound scan Created 2026-10-20 Updated 2026-11-05
An M-mode ultrasound scan repeatedly samples one pulse-echo ultrasound beam and places successive depth profiles alongside one another in time. Depth is one display axis and acquisition time the other; echo strength controls brightness. This makes motion along the selected line visible. [ref]
A stationary interface traces a line at constant depth; an oscillating interface traces a varying depth. Converting echo delay to depth still assumes a sound speed and a return path. This display records one beam over time, whereas a B-mode ultrasound scan combines beam lines across space.