A laboratory <vacuum> is a region with substantially reduced gas <pressure>, rather than an experimentally perfect absence of <matter>. Lowering the <number density> increases the <mean free path>, allowing a <molecular beam> to travel with fewer <collisions>. In <molecular flow>, the <mean free path> exceeds the apparatus dimensions, so wall encounters dominate.
A <vacuum pump> competes with leaks and <outgassing>. For an isothermal vessel of <volume (thermodynamics)> $V$, constant <gas load> $Q$ in pressure-volume per unit <time>, and effective <pumping speed> $S$, gas balance gives
$$
V\frac{dp}{dt}=Q-Sp,\qquad p(t)=\frac QS+\left(p(0)-\frac QS\right)e^{-St/V}.
$$
This ideal model predicts a pump-down <time constant> $V/S$ and a limiting <pressure> $Q/S$. A restricted connecting pipe reduces the effective <pumping speed>; a larger pump alone need not fix it. Measuring the <pressure> transient tests the model, while deviations can reveal changing <outgassing> or <pumping speed>.<Vacuum Technology for Ion Sources by Paolo Chiggiato>[sections 2.2–2.5 and 3.1–3.3: flow regimes, conductance, gas release and pressure transients]{ref}
\Image[/vacuum-pump-down.png]
{title=Ideal pump-down curves show an exponential transient and a floor set by the gas load}
{description=Original calculations with stipulated constant volume, effective pumping speed and gas load. These illustrate the isothermal balance equation, not measured pump performance.}