A vector network analyzer measures complex reflection and transmission of electromagnetic signals as a function of frequency. The word vector records that both magnitude and relative phase are measured. It is used in electrical measurement of devices such as transmission lines, filters and amplifiers; the measurement is a comparison of outgoing signals with a reference excitation.
A primary manufacturer explanation describes a tuned receiver that down-converts the measured signal and compares it with a reference split from the source. Its test set samples the reference signal automatically. A longer signal path produces a frequency-dependent phase delay, even if its magnitude is unchanged.[ref]
For a two-port device, scattering parameters relate incident wave amplitudes to outgoing amplitudes :
When port 2 has no incident wave, and . The termination and reference-plane conditions belong to the definition. For equal real reference impedances with power-normalized waves, is the transmitted-to-incident power ratio under that condition.
For example, has magnitude , phase and power ratio . Its logarithmic magnitude is . A magnitude-only measurement loses the phase. A pure delay contributes ; a delay changes phase by over a frequency increase, without changing magnitude. This is an original numerical example of the manufacturer's reference-delay principle.
Instrument calibration and a stated reference plane are consequently important: connecting cables and fixtures can contribute their own response. Phase wraps also require care because angles differing by describe the same complex value. The complex relative gain error can quantify a deviation that would be invisible if only transmitted magnitude were compared.