A researcher in magnetic measurement whose account of demagnetizing factors distinguishes the field applied to a specimen from the field produced by the specimen’s own magnetization. For a cylinder, that internal field is generally nonuniform, so an instrument sampling the central region and one averaging the entire specimen need not require the same correction. This explains why specimen shape and measurement geometry matter when inferring a material’s magnetic susceptibility.[ref]
Demagnetizing Factors. Printed p. 104, §2: fluxmetric and magnetometric definitions, cylindrical nonuniformity and susceptibility dependence