XY8 is an eight-pulse refocusing phase pattern whose nominal rotations follow
Here and denote rotations about perpendicular axes of the rotating frame, not different flip angles. The pattern is used to reduce cumulative NMR pulse errors and is also used in broader spin-control experiments.[ref]
The 1990 compensated-sequence paper compared x/y phase-alternating pulse trains experimentally and described compensation of errors affecting all three magnetization components, including applications to REDOR. The benefit concerns imperfect pulse trains: ideal rotations alone would not demonstrate that real pulse errors cancel.[ref]
For an original idealized check on a magnetization vector, a rotation about has matrix , while . Multiplying the eight matrices in the stated order gives the identity. All three vector components therefore return to their initial values if free evolution and pulse imperfections are omitted. This check verifies the ideal net action; it does not model the cancellation of imperfect rotations.
Three repeated XY8 blocks contain24 refocusing pulses, before counting preparation and readout pulses. With instantaneous pulses separated by a centre-to-centre interval and half-interval end gaps, their ideal evolution window is . Finite pulse durations and the actual interaction being refocused must be included in a quantitative experiment.
Jacob Schaefer's retrospective reports that applying XY8 to REDOR reduced off-resonance amplitude and phase distortions in long pulse trains, enabling measurement of weak NMR dipolar couplings. That is a concrete experimental motivation for phase routing, rather than a claim that XY8 removes every kind of noise.[ref]