Publisher: National Institute of Standards and Technology.
Cited passages: A New NIST Automated Calibration System for Industrial-Grade Platinum Resistance Thermometers, §§3.1, 3.3 and 4.2, printed pp. 2–6: baths, readout, four-terminal devices and reversed current; §1, printed p. 2; §4.2, printed pp. 5–6; §5.4, printed p. 9: excluded effects, drift rejection, check thermometer, immersion and extra fit checks.
NISTIR 6225, published 1 September 1998.
Experimental journal article published in 1997 and credited in print to G. T. Furukawa, B. W. Mangum and G. F. Strouse. The full names correspond to George T. Furukawa, Billy W. Mangum and Gregory F. Strouse. NIST hosts the article and its publication record.
Cited passages: Furukawa, Mangum and Strouse, Effects of different methods of preparation of ice mantles of triple point of water cells on the temporal behaviour of the triple-point temperatures (1997), Abstract, §1 and §§3.2–3.5: aged reference, inner melt, bridge readout, zero-power and hydrostatic corrections.