Platinum resistance thermometer
ID: platinum-resistance-thermometer
A resistance thermometer using platinum as its sensing material. It provides the temperature readout in the saturation-pressure equilibrium measurement.
Transduction turns electrical resistance into a temperature indication. A uniform wire has , with electrical resistivity , length , and cross-sectional area . Stronger atomic vibrations at higher temperature scatter Electrons more, increasing platinum's electrical resistivity; impurities and crystallographic defects also contribute.[ref]
Guide on Secondary Thermometry: Industrial Platinum Resistance Thermometers by Jonathan Pearce and others. Guide on Secondary Thermometry: Industrial Platinum Resistance Thermometers, §2.1, printed pp. 6–7: resistivity, electron scattering, and impurities
A small known electric current and measured electric potential difference give through Ohm's law; a resistance bridge can instead compare the element with a reference resistor. Inverting the calibrated response then gives the sensing element's temperature. This makes electrical readout convenient, but the platinum resistance thermometer lead resistance error, resistance thermometer self-heating, and resistance thermometer immersion test explain why a stable display alone is insufficient. The gas thermometer supplies a contrasting route through an equation of state; the saturation-pressure equilibrium measurement shows why a trustworthy independent temperature readout matters.
New to topics? Read the docs here!