In piezoresponse force microscopy, a conducting atomic force microscopy tip contacts a sample with a counter-electrode. Apply . The converse piezoelectric effect produces an oscillatory displacement, approximately . Contact with the moving surface bends the cantilever, and a lock-in amplifier records the displacement amplitude and phase relative to the drive.
For opposite out-of-plane ferroelectric domains, the effective piezoelectric coefficient changes sign. Their vertical response therefore differs by approximately in phase. Equivalent antiparallel domains can have similar amplitudes, while cancellation or reduced response near a wall can produce an amplitude minimum. Phase identifies the relative response sign; an uncalibrated detector phase is not an absolute polarization direction.
Use a small drive below the switching threshold while imaging, a stable contact force and a well-characterized contact resonance. Electrostatic forces, contact stiffness and surface charging can also create signals at the drive frequency; observing contrast alone is insufficient evidence for polarization switching.[ref]
Imaging mechanism of piezoresponse force microscopy of ferroelectric surfaces by Sergei V. Kalinin and Dawn A. Bonnell. 2002, electromechanical and electrostatic contrast regimes
Piezoresponse force microscopy 2026-10-20
A contact measurement of electrically driven local surface displacement, detected by cantilever bending or torsion and often demodulated by a lock-in amplifier. Electrostatic and contact-mechanical cross-talk must be separated from converse piezoelectric effect response.