A detection decision threshold is a cutoff applied to an observed analytical response to decide whether there is evidence for an analyte contribution beyond background. A decision above the cutoff can be wrong even for an analytical blank: this is a type I error. Missing a genuine contribution is a type II error. The chosen rule and the measurement procedure determine those error probabilities.[ref]
For an original one-sided example, suppose blank responses have a known normal distribution . Set . Declaring detection when gives about a5percent type I error probability. A real sample whose mean is exactly has only a50percent chance of crossing that same detection decision threshold.
Under the additional assumption of unchanged standard deviation, a mean near gives about95percent probability of crossing it: the mean lies another units above . Thus an observed-response cutoff and a limit of detection giving a specified detection probability answer different questions. These numbers are a known-variance teaching model, not a universal laboratory convention. Estimating the background, changing the error criterion or allowing signal-dependent variance changes the calculation. The threshold must be tied to the complete procedure whose detection performance is claimed.[ref]
Lloyd A. Currie 2026-11-05
Lloyd A. Currie prepared the 1995 IUPAC recommendations on evaluating analytical methods, including detection and quantification capabilities. A central contribution was to separate an observed-result decision from the capability of a measurement process. A detection decision threshold controls the chance of a false positive, while a limit of detection also depends on the chance of missing a real signal.[ref]
An original idealized example makes the distinction visible. Suppose the blank-corrected measurement has a normal distribution with standard deviation one. A threshold near gives a 5% type I error under a zero-mean blank model. A true mean near places that same threshold about standard deviations below the mean, giving about 95% detection probability, or 5% type II error.
The two signal levels answer different questions: whether one observed result crosses a decision boundary, and how strong a true signal must be for reliable detection under the assumed process. These numbers belong to the stated model; they are not universal detection limits for instruments.