Prepare a dilute dispersion of separated nanoparticles on an electron-transparent support, for example a thin carbon film on an electron-microscopy grid. A sparse deposited aliquot can be dried and the grid mounted over an open transmission aperture in the holder. The beam path must not be blocked by a bulk substrate.
Dilution and a thin, clean support reduce overlap, multiple scattering and background. Remove excess soluble residue where compatible with the particle surface chemistry, and ensure the support and mounting provide adequate electrical conduction. The desired object is an isolated, vacuum-compatible particle layer, rather than a thick dried droplet or an opaque slab.
Use a sufficiently high SEM acceleration voltage to transmit through the particle/support combination and provide useful diffracted intensity; operation near the high-voltage end of the SEM range is a reasonable starting point. The supplied wavelength is used directly for the camera-length calculation, rather than assigning a precise voltage from an approximate wavelength relation.
Choose the smallest beam current consistent with usable counting statistics, because the stated source-size contribution obeysHigher gun brightness permits more current at the same small probe. A relatively short working distance can give a larger convergence semiangle and reduce this term. In the specified fixed lens-to-detector geometry, shortening working distance also increases the specimen-to-detector camera length, reducing the detector's collection angles.
The objective aperture should be optimized, not made arbitrarily large: increasing reduces the source-size and diffraction contributions, but spherical/chromatic aberrations and angular broadening eventually worsen the probe or diffraction selectivity. A suitable aperture balances those effects while keeping the chosen diffraction rings within the annulus. Adjust specimen height to meet that angular condition, focus and align at the operating voltage, and control drift and contamination.
Low convergence favors angular discrimination of diffraction, whereas high convergence can favor probe size. The optimum therefore depends on whether spatial detail or sharply selected Bragg diffraction contrast is the limiting requirement.