In photoemission electron microscopy, illuminate a surface with ultraviolet light or X-rays and form an electron-optical image of the emitted electrons. An extraction field accelerates them away from the sample; an immersion objective and projection lenses magnify their spatial distribution onto an electron detector. Unlike a scanned probe, this is a full-field image. Surface preparation and vacuum matter because the emitted low-energy electrons are strongly affected by contamination and local electric fields.
Contrast can come from work-function differences, composition, occupied electronic states, illumination and topography. Near threshold, a small work-function change can produce a large emission-yield change. Therefore brightness is not simply geometrical height. An energy filter can select an emitted-electron energy interval; scanning the selection yields spatially resolved spectra. Sweeping the photon energy across an absorption edge can add element and chemical-state sensitivity. Appropriate polarization-dependent measurements can probe magnetic domains.[ref]
Photo emission electron microscope by JEOL. instrument definition and optional energy-filter imaging
Resolution depends on electron-lens aberrations, the accepted angular and energy spread, extraction geometry, sample flatness, space charge, mechanical stability and detector sampling. Rejecting electrons improves some aberrations at a cost in signal. An unfiltered photoelectron image and a chemically selected image therefore answer different questions even when they use the same microscope.