Polarity reversal measurement separates a signal that changes sign on reversal from an additive contribution that does not. If readings in two polarities obeythen the half-difference estimates the signal, , while the half-sum estimates the unchanged offset, . The procedure depends on reversing the intended signal while preserving the relevant offset.
A primary electrical application is cancellation of thermoelectric offset in low-level measurements. The Keithley handbook describes reversing sources when comparing small electric potential differences, and reversing test current for low-electrical resistance measurements. A thermal offset can arise at dissimilar-metal junctions and remain approximately unchanged during a sufficiently short cycle.[ref]
For an original resistance example, a bipolar current source applies and to a resistor while a voltmeter measuresLet and readings be and . Their half-difference is , giving ; their half-sum reveals a offset. Using only the positive reading would instead give .
An offset changing from to leaves resistance error . A change in this example therefore leaves error. Allow electrical settling after reversal while keeping the cycle short relative to material temperature changes. The handbook explicitly notes this time-scale requirement.[ref]
Low Level Measurements Handbook, Seventh Edition by Keithley Instruments. §3.3.2, current-reversal method, thermal-time-constant limitation
Reversal does not automatically remove lead resistance, current-magnitude errors, self-heating or every offset in the circuit. Record the actual reversal operation and timing with the measurement procedure. Articles by others on the same topic