A null detector indicates whether a residual signal has been brought sufficiently close to zero in a null measurement method. A galvanometer between the middle junctions of a Wheatstone bridge is a concrete example: equal junction potentials give zero detector current.[ref] Its useful property is the ability to distinguish small residuals and their sign near balance.
For an original electronic model, let the detector output be
where is the residual electric potential difference, is gain, an output zero offset and an output noise term. With known zero offset and negligible noise, an output threshold at corresponds to . It is a resolution criterion for this model, not a total measurement uncertainty.
A false zero exposes another limit. If , a displayed can occur at for the same gain. Increasing gain improves the output response to a given residual, but measurement calibration must still address offset and noise. Polarity reversal measurement can help separate signal from an unchanged internal offset: if
then and . The separation assumes reversal changes the signal sign while leaving the offset unchanged.
Detector loading also matters away from null. For a residual source with series electrical resistance connected to detector input resistance , the detector receives . The ideal zero stays at , but a low input resistance reduces the response to a small departure from it. Verify the response with known small positive and negative signals, allow settling, and investigate persistent offsets. A quiet zero may otherwise reflect a broken connection or insufficient sensitivity rather than actual balance. Check the working range when approaching null because detector saturation invalidates the assumed linear response.

Articles by others on the same topic (0)

There are currently no matching articles.