Gel permeation chromatography separates by solution hydrodynamic volume, rather than molar mass alone. At equal molar mass, a more extended chain accesses fewer pores and elutes earlier. The relatively rigid, conjugated poly(3-hexylthiophene) is the natural first-eluting candidate, compared with the more flexible chains, provided all samples are molecularly dissolved.
This is a structural prediction, not an unconditional ordering: solvent quality, polymer aggregation and column interactions affect retention, and the hydrogen-bonding polyamide may not dissolve adequately in tetrahydrofuran. An insoluble sample cannot be assigned an ordinary solution gel permeation chromatography elution time.
A differential refractive index detector can detect all four polymers when their refractive-index contrasts are nonzero. Light scattering with an appropriate concentration detector can additionally provide molar mass. A single ultraviolet detector is unsuitable as the universal detector because polypropylene lacks the strong chromophores present in the aromatic and conjugated products.
Refractive index detector Created 2026-10-20 Updated 2026-11-05
A refractive index detector measures the difference in refractive index between a sample stream and a reference. In liquid chromatography, a solute passing through the detector can change that difference and produce a signal.
The signal depends on both solute concentration and the solute's refractive-index increment relative to the solvent. Temperature drift or changing mobile-phase composition can also shift the baseline, so a peak is not automatically a substance-specific identification.