Spectral leakage bias in harmonic distortion measurements occurs when a finite record spreads a strong component into bins interpreted as distinct harmonics. A waveform can have zero physical total harmonic distortion while a naive FFT-bin ratio reports otherwise. Confining a sinusoid to a finite interval creates spectral sidelobes; the pinned primary guide derives this leakage from the finite-time Fourier transform.[ref]
For an exact discrete calculation, take , where is the number of cycles in the record. With the unscaled forward discrete Fourier transform, a finite geometric sum gives
where
whenever the denominator is nonzero, with the finite-sum limit elsewhere. Integer-cycle coherent sampling makes the unwanted integer bins cancel; a noninteger cycle count generally does not. The negative-frequency contribution also matters because the input is real.
For an original pure-tone example, let , , and . Its true higher-harmonic amplitudes are zero. Yet choosing the nearest bins , , and for the fundamental, second and third harmonic gives
A naive result would be about distortion from this undistorted sinusoid. This is an exactly specified record calculation, not an instrument's measured distortion floor.
A window reduces some sidelobes at the cost of a broader main lobe and a changed amplitude normalization. Its coherent gain of a window must be treated consistently, and one cannot assume that one selected bin captures a noncoherent tone's full amplitude. Zero padding samples the same finite-record spectrum more densely; it does not restore missing observation time or remove leakage.
A measurement can instead synchronize the source and acquisition for an integer number of cycles, fit a justified sinusoidal model, or use a validated window/band-integration method. The finite-record phasor extraction with coherent sampling gives a particularly transparent exact case. Calibrated receiver response and actual harmonic coverage remain necessary, as in the primary Anritsu apparatus.[ref] Power spectral density and squared-amplitude spectrum outputs also have different units; an arbitrary square root of a density bin is not a calibrated harmonic amplitude.[ref]