Lateral position jitter
ID: lateral-position-jitter
Lateral position jitter is the fluctuating in-plane displacement of successive acquired images relative to a chosen reference. In serial block-face scanning electron microscopy, it can displace corresponding features between block faces even when the cutting increment is perfectly regular. A simple image registration model iswhere is the ideal image of the current face and is its lateral shift. The specimen itself changes between faces, so not every apparent shift is instrumental motion.
A primary apparatus study estimated inter-image displacement from cross-correlation peaks in several subregions. Its Figure4 central cluster had standard deviations of and along the two axes; outliers and changing specimen patterns required interpretation.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Results, registration paragraph; Figure4 and caption; Materials and Methods, Data analysis
For an original one-axis example, take measured frame offsets pixels at pixel pitch . The offsets are and their root mean square about zero isThe successive differences are , whose root mean square is . These are different statistics. For independent equal-variance frame offsets, the theoretical difference variance is ; correlation or drift changes that relation. Specify the reference, detrending and normalization when reporting jitter. Image registration can estimate a correction but does not prove that all residual motion or specimen deformation has been removed.
New to topics? Read the docs here!