Serial block-face scanning electron microscopy acquires a three-dimensional dataset by repeatedly imaging an exposed specimen face in a scanning electron microscope and removing a thin layer with an ultramicrotome. The removed layer is not the imaged object: each image records the newly exposed block face. This differs from collecting and imaging thin sections in serial-section electron microscopy.[ref]
Serial Block Face Scanning Electron Microscopy by JEOL. Definition, in-chamber sectioning and imaging of the newly exposed face
In the 2004 apparatus of Winfried Denk and Heinz Horstmann, a diamond knife operated inside the microscope chamber, with controlled specimen advance and retraction. A backscattered electron signal supplied image contrast. Keeping the cutting and imaging operations in the chamber enabled automated image stack acquisition.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Results, Block-Face Imaging; Materials and Methods, Microtome
For an original ideal geometry, acquire the first face at , then advance and remove before each subsequent image. An image stack of faces has its last acquired face atThis requires 100 inter-image cuts. If a data format assigns each face a -deep voxel, its nominal represented thickness is instead . State which extent is intended. At lateral pixel pitch , the sampling ratio is 8.
The resulting image stack needs image registration and a calibrated cutting coordinate. Lateral position jitter, cutting artifacts and the depth range contributing to the electron signal limit the useful reconstruction. Finite voxel spacing is not itself a measurement of spatial resolution. Destructive cutting also means that later imaging cannot recover an exposed face that has already been removed.
Lateral position jitter is the fluctuating in-plane displacement of successive acquired images relative to a chosen reference. In serial block-face scanning electron microscopy, it can displace corresponding features between block faces even when the cutting increment is perfectly regular. A simple image registration model iswhere is the ideal image of the current face and is its lateral shift. The specimen itself changes between faces, so not every apparent shift is instrumental motion.
A primary apparatus study estimated inter-image displacement from cross-correlation peaks in several subregions. Its Figure4 central cluster had standard deviations of and along the two axes; outliers and changing specimen patterns required interpretation.[ref]
Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure by Winfried Denk and Heinz Horstmann. Results, registration paragraph; Figure4 and caption; Materials and Methods, Data analysis
For an original one-axis example, take measured frame offsets pixels at pixel pitch . The offsets are and their root mean square about zero isThe successive differences are , whose root mean square is . These are different statistics. For independent equal-variance frame offsets, the theoretical difference variance is ; correlation or drift changes that relation. Specify the reference, detrending and normalization when reporting jitter. Image registration can estimate a correction but does not prove that all residual motion or specimen deformation has been removed.