A laboratory micropipette is a pipette for transferring small measured quantities of liquid, commonly expressed in microlitres. In piston-operated models, the displacement mechanism matters. An air-displacement pipette leaves an air cushion between piston and sample, whereas a positive-displacement pipette places the piston directly against the liquid.[ref]
Weighing a delivered sample provides a concrete check. In a simplified calculation that neglects air buoyancy and evaporation, liquid volume follows from mass and density :For an assumed , a delivered mass of corresponds to . Relative to a nominal setting, the deviation is . These are illustrative values, not calibration data.
A repeated series can distinguish a displaced mean from wide scatter: measurement accuracy and standard deviation describe different aspects of performance. Actual gravimetric checks account for temperature-dependent water density, air buoyancy and evaporation losses. The manufacturer's method uses a conversion factor and an evaporation correction, showing why a displayed volume alone does not establish the quantity delivered.[ref]
Gilson Guide to Pipetting. pp. 46–49, systematic/random error, gravimetric conversion and repeated weighings