Sinusoidal-stage optical trap calibration
ID: sinusoidal-stage-optical-trap-calibration
For an overdamped harmonic optical trap, impose a known sinusoidal local fluid velocity of amplitude and angular frequency . The coherent bead response has amplitude . Measuring the thermal corner frequency and detector-voltage amplitude therefore determines the position sensitivity in the displayed formula without a drag coefficient as input. Use a thermal record with the drive line removed to obtain stiffness from the equipartition theorem. Fluid-following, detector response and hydrodynamic corrections must remain valid. Combined thermal and stage measurements have experimentally calibrated an optical trap without drag as an input.[ref]
Calibration of optical tweezers with positional detection in the back-focal-plane by Simon F. Tolić-Nørrelykke and others. 2006 preprint abstract: calibration without drag as an input