Third-order correlation brightness weighting

ID: third-order-correlation-brightness-weighting

Third-order correlation brightness weighting is the way bright shots contribute disproportionately to an averaged nonlinear correlation trace. A normalized average trace is a weighted mean of shot-normalized shapes, with weights set by their central correlation signals rather than by how often the shots occur.
Let a nonnegative, background-corrected third-order trace on shot be . Define its central signal and normalized shape . Then the experimentally distinct averages are
and the equal-shot mean . The nonnegative sum to one, but are equal only when the are equal. Here “central” denotes the specified zero-delay signal, not whichever noisy sample happens to be the largest.
For identical pulse shapes multiplied by intensity factors , the ideal integral scales as . The primary correlator implements this three-intensity-factor gate using second-harmonic generation followed by sum-frequency generation.[ref] If shapes change, can depend on duration and overlap as well as brightness, so pulse energy alone need not determine the weight.
An original two-class example has equally frequent classes with and at one chosen delay. Their equal-shot mean ratio is , but the normalized summed trace gives
The bright class supplies eight ninths of the central signal despite representing only half the shots. Neither result is an arithmetic mistake: they estimate different summaries of the stipulated ensemble.
For an ensemble with finite moments and , the distinction can be expressed exactly through covariance:
Unchanged normalized shape makes the covariance zero, even with large brightness fluctuations. Shape–brightness dependence can make it nonzero; cubic intensity-noise bias in pulse correlation alone does not prove a contrast change.
A shot-resolved pulse-contrast estimator needs synchronized intensity information and a stated selection rule. Thresholding weak shots or dividing by a noisy central signal can introduce further selection and ratio errors. Report whether normalization occurred before or after averaging, retain detector baseline subtraction, and test readout linearity. A smooth averaged trace is not a representative single shot unless those assumptions are justified.