A microplate is a tray containing an ordered array of small wells that keep samples physically separate while permitting repeatable handling and measurement. A microplate reader addresses wells to measure optical signals. The plate is consequently part of the measurement geometry: its positioning, well shape and material affect how illumination reaches the solution and how transmitted or emitted light is collected.
The microplate footprint is the base outline used to fit the plate into handling equipment. The primary footprint standard specifies nominal length127.76mm and width85.48mm. Its tolerances distinguish measurements within12.7mm of the corners, , from measurements anywhere along the sides, . The stated dimensions apply at unless otherwise specified.[ref] Nominal dimensions alone therefore do not define the entire fit envelope.
For an original counting and volume example, an8-row,12-column layout has96 wells. Filling each with200µL requires19.2mL of delivered liquid, before losses in transfer equipment are included. Equal delivered volume does not itself prove equal concentration, and a plate-wide mean can conceal an individual underfilled well. Keeping the row/column address with each reading allows sample, reference and repeat locations to remain distinguishable.
For an ideal flat-bottom well, a volume estimate can be related to depth by , where is its internal liquid-covered area. A hypothetical and give5mm depth. Curved bottoms, taper and the meniscus alter that simple geometry. In transmission measurements the resulting depth enters the Beer-Lambert law, making microplate path-length correction relevant to comparisons across filling volumes.
The standard's foreword records dimensional variation across suppliers as a source of problems in automated instruments and describes standardization work beginning in1995.[ref] A common footprint promotes mechanical compatibility; it does not imply identical well volumes, optical backgrounds or spectral transmission. Those properties still need to match the intended physical measurement.