AFM tip convolution 2026-10-20
Geometrical broadening of a surface feature by the finite shape of an AFM probe. The measured height envelope is more precisely a morphological dilation than an ordinary linear convolution. Repeated identical particle outlines can be tip images; inaccessible sidewalls prevent a unique inverse reconstruction.
In tapping-mode atomic force microscopy, demodulate the cantilever displacement to obtain its amplitude and compare it with a setpoint below . A closed-loop control system drives a piezoelectric scanner in to restore the setpoint as the raster crosses raised or depressed regions. The interaction remains in its sensitive amplitude range while the scanner follows heights much larger than the oscillation amplitude. The principal height image is the calibrated feedback displacement at each lateral pixel. The amplitude-error and phase channels are supplementary signals, rather than interchangeable height measurements. Maximum height is now bounded by scanner travel, clearance and stable tracking. Steep features require adequate feedback bandwidth and sufficiently slow scanning; trace/retrace disagreement tests tracking, while a changed AFM probe tests geometric broadening. Surface-dependent dissipation can move the setpoint height even on a geometrically flat specimen.
A steady-current image is not supported by the electrical connection drawn. tunnelling atomic force microscopy requires a conducting AFM probe, an applied bias voltage and a continuous return path from the contacted region to the current amplifier. graphene is conducting, but a flake on insulating glass is not electrically grounded merely because the other side or edge of the glass is earthed. Charging transients or displacement current are not its steady conductivity. Add a metal contact actually touching the graphene before using its current contrast; ordinary force-based surface topography can still be recorded without that contact. The manufacturer's nanostructure measurements explicitly distinguish objects connected to conducting pads from unconnected particles on the same insulating substrate.[ref]
Simultaneous Electrical and Mechanical Property Mapping at the Nanoscale with PeakForce TUNA by Bruker. application note 132, Figure 11 and accompanying discussion of electrical connection
The repeated, similarly oriented rounded triangular outlines in panel (a) are characteristic of AFM tip convolution: a narrow raised object can image the shape of a blunt or contaminated AFM probe, rather than reveal its own lateral shape. A finite tip radius broadens lateral dimensions. Replace or clean the AFM probe, check it on an independently characterized sharp-feature standard, and repeat at low load; agreement after a tip change is stronger evidence than a visually pleasing image. Image processing cannot uniquely recover unmeasured sidewalls or prove that every particle with a similar outline is an artefact.
Panel (b) contains horizontal streaks associated with steep features. The natural instrumental diagnosis is AFM feedback artefact: the height controller does not track a rapid change faithfully, or rings after it. Reduce the scan speed and adjust proportional/integral gains below the oscillation threshold. Compare trace and retrace and rotate the fast-scan direction: a streak following the scan rather than the specimen is instrumental. Loose contamination can also produce scan-line streaks, so clean or change the AFM probe if tuning does not remove them. Tip-shape broadening in (a), and scan-direction tracking/streaking in (b), are the principal diagnoses; the images alone do not uniquely establish every underlying cause.
At position (1), the AFM probe is far from the surface, with approximately zero force. At (2), attractive interactions bend the AFM cantilever toward the surface; when the attractive force gradient exceeds the cantilever spring constant, the equilibrium becomes unstable and jump-to-contact occurs. Position (3) is on the mechanical-contact branch: changes in imposed height largely bend the AFM cantilever and, for a compliant specimen, indent the sample. This branch can include compressive contact and adhesive contact under tension; contact does not itself imply a positive force at every point. At (4), during withdrawal the tip remains attached and pulls the AFM cantilever downward until the pull-off force is reached and it detaches. Approach and retraction follow different paths because adhesion creates a hysteretic contact. The numbered points identify stages on the source curve, not four extra question parts.