A balloon compliance fixture correction removes the volume response of tubing, connectors, fluid and other apparatus from a measured balloon volumetric compliance. It is a correction to a defined measurement arrangement. The pressure reference and the state of the apparatus must match between characterization and use.
For a quasistatic experiment in which fixture and balloon experience the same pressure increment, write
A separately characterized fixture supplies . A rigid termination can provide one control configuration, but changing tubing or trapped gas between the control and balloon runs changes the correction itself.
For an original example, and give . Assigning the whole measured response to the balloon would overestimate it by one third relative to the corrected value. Independent standard uncertainties and give . Correlated inputs instead require the term in the variance.[ref]
Trapped gas explains one nonlinear contribution. For an isothermal gas pocket with constant, its local stored-volume response is . A pocket at absolute pressure contributes . Absolute pressure is required here; substituting gauge pressure into the gas law would be wrong.
A manufacturer's diameter-versus-pressure chart, such as Nipro's Thouzer chart, reports a different observable from injected volume.[ref] Such a chart does not measure the fixture's stored volume. Comparing a measured volume curve with it requires the balloon geometry and apparatus correction, rather than treating diameter compliance and whole-system volume compliance as identical.
During flow, different parts of the fixture can have different pressures. The simple additive quasistatic model then needs a dynamic extension, consistent with the balloon inflation hydraulic time constant. Loading history and hysteresis must also match before two response curves can be subtracted.