Branch-current mismatch in a resistance ratio measurement is a residual scale error when the sensor and reference resistor carry unequal electric currents. A common supply or common nominal current setting does not establish the shared excitation required by a shared-excitation resistance ratio measurement.
Suppose separate branches carry positive currents and . Ideal potential sensing givesThus the fractional resistance error is exactly . If and , the common scale cancels, whileIn an unbranched series circuit the same electric current passes through both components; this particular error instead concerns separate paths, leakage, or unequal effective sampling times.
For an original example, use , , and . Their potentials are and . The inferred resistance is , high by approximately . Each current is only from its nominal setting, but the opposed errors add to first order.
A useful ideal comparison is current-source swapping in a resistance ratio measurement. In two otherwise unchanged acquisitions, assign source A to the sensor and source B to the reference, then exchange them. If , the two ratios obeyThis exact geometric mean cancellation assumes unchanged source outputs and resistances between acquisitions; switching transients and sensor self-heating in a ratiometric resistance measurement can violate that assumption. The primary circuit guide independently treats current-source mismatch and source interchange in its three-wire lead-cancellation circuit; that topology's correction must follow its own equations rather than be replaced by the two-branch model above.[ref]
A proposed test measures both branch currents through characterized sensing resistors, then changes their assignments while holding the sensor environment fixed. Agreement after swapping tests the mismatch model; agreement from merely changing the common nominal current tests a different cancellation.
Current-source swapping in a resistance ratio measurement exchanges two excitation sources between the sensor and reference resistor, allowing an unequal-current contribution to be identified or canceled. The combining rule depends on the circuit topology. In the separate-branch model of branch-current mismatch in a resistance ratio measurement, the exact combination is a geometric mean.
Let and let the two positive source currents have a fixed ratio . With source A on the sensor and B on the reference, the measured potential ratio is . After exchanging their assignments, . HenceThis assumes that the resistances, current outputs, potential-sensing gains and relevant thermal conditions are unchanged across the two acquisitions. It cancels the multiplicative source ratio, not an additive readout error or an inaccurate assigned reference resistance.
For an original example, let and . The acquisitions give and . Their geometric mean is exactly . Their arithmetic mean is , high by approximately . In general,The arithmetic mean suppresses a small mismatch to second order here, but is not the exact inverse-pair cancellation.
The primary RTD guide uses source interchange in a different three-wire lead-cancellation circuit. Its two readings are averaged according to that circuit's equations; they must not be replaced automatically by the geometric-mean rule above.[ref]
A drift residual after current-source swapping remains if the source ratios in the two acquisitions are and . For unchanged , the same calculation gives . A increase from to therefore makes the combined estimate low by . Sensor self-heating in a ratiometric resistance measurement can also change between source assignments.
A proposed apparatus check switches two characterized current sources between stable sensing resistors, waits for electrical settling, and compares both individual readings and their product. Reversing the order of the two acquisitions helps distinguish a repeatable source mismatch from drift during the sequence. This tests the stated circuit model without presuming that all source-swapping circuits use the same estimator.