Four-terminal sensing uses one pair of contacts to carry electric current and a separate pair to sense electric potential difference. The sensing instrument draws negligible electric current, so the voltage drops along its own leads are negligible, and the current-contact drops outside the sensing interval are excluded. A measurement using just two shared leads includes their electrical resistance and contact drops, which can overwhelm the small sample signal in a resistance-versus-temperature experiment for superconductivity. Four contacts do not automatically remove offsets from the thermoelectric effect. [ref]
https://www.tek.com/en/documents/technical-article/low-voltage-measurement-techniques. Tektronix, Low-Voltage Measurement Techniques, four-lead arrangement and thermoelectric offsets
For a stable offset , let denote the magnitude of the reversed electric current. The measured electric potential differences are and , so subtraction gives the displayed electrical resistance. For example, with stipulated , and , the two readings are and ; their difference correctly recovers . If the offset changes between readings, its change leaves a residual measurement error; the sample and measuring circuit must remain sufficiently stable during reversal.