Ahmet Baydur 2026-11-05
Ahmet Baydur is first author of the 1982 primary study testing an esophageal balloon method for estimating pleural pressure. His official faculty profile identifies respiratory mechanics and control of ventilation among his research interests.[ref]
A Simple Method for Assessing the Validity of the Esophageal Balloon Technique by Ahmet Baydur and others. Primary byline; abstract, experimental objective
[ref]The esophageal balloon occlusion test experiment occluded the airway at end expiration in ten subjects and compared changes of esophageal and mouth pressure during ensuing inspiratory efforts. It tested sitting, supine and lateral positions. The reported pressure-change ratio was near unity in sitting and lateral positions; some supine deviations could be reduced by repositioning the balloon. This made balloon placement and posture testable parts of the measurement rather than assuming that any esophageal pressure trace was an adequate surrogate.[ref]
A Simple Method for Assessing the Validity of the Esophageal Balloon Technique by Ahmet Baydur and others. Abstract, occlusion protocol, posture comparison and repositioning observations
For an original reduction, suppose an occluded effort changes mouth pressure by cm of water and esophageal pressure by cm of water. The change ratio is . If repositioning gives cm of water at the same mouth change, the ratio becomes 0.98. These invented numbers illustrate the comparison; they are not observations from the paper or a prescribed acceptance threshold.
A change ratio also cannot determine a constant offset. If , then for fixed . Agreement of changes therefore addresses pressure transmission under the tested conditions, rather than proving equality of every absolute pressure reading.
Ahmet Baydur, identified portrait linked from his USC faculty profile
. Esophageal balloon catheter 2026-11-05
An esophageal balloon catheter is the tube-and-balloon instrument used to transmit a pressure signal from a balloon in the oesophagus to a pressure sensor. The esophageal balloon is the coupling element; the catheter also introduces a gas-filled line, connections and an external measurement chain. Those components can affect both baseline pressure and dynamic response.
The physiological filling study by Francesco Mojoli and collaborators shows why treating the assembly as an ideal pressure wire is inadequate: changing balloon filling altered pressure-change transmission and wall-related baseline pressure.[ref]
In vivo calibration of esophageal pressure in the mechanically ventilated patient makes measurements reliable by Francesco Mojoli and others. Abstract, Methods and Results
This establishes a measurement dependence, not one universal filling volume for every catheter and surrounding pressure.A minimal model of pneumatic response lag of an esophageal catheter iswhere is a varying source pressure and the sensor output. For a step from to , the response isWith a stipulated , 63.2 percent of the step appears after 0.10 s and 95.0 percent after 0.30 s. These are model calculations, not specifications of a clinical device.
This dynamic error differs from a fixed offset: waiting longer removes the modeled step-response lag but not a baseline bias. A comparison against a known pressure waveform can test the instrument's response, while an esophageal balloon occlusion test addresses transmission in the measurement setting. Neither validation alone proves that esophageal pressure equals regional pleural pressure throughout the lung.
Esophageal pressure measurement 2026-11-05
Esophageal pressure measurement records pressure in the oesophagus, commonly with an esophageal balloon, to obtain information about respiratory mechanics and pleural pressure. The measured signal is a local surrogate. It is not a direct map of pressure everywhere in the pleural cavity.
A primary validation experiment by Ahmet Baydur and collaborators compared changes in esophageal and mouth pressure during end-expiratory airway occlusion and spontaneous efforts. Ten subjects were studied in different postures; some discrepancies changed when the balloon was repositioned.[ref]
A Simple Method for Assessing the Validity of the Esophageal Balloon Technique by Ahmet Baydur and others. Abstract: subjects, posture, occlusion and repositioning
The esophageal balloon occlusion test therefore evaluates pressure-change transmission under specified conditions, rather than proving all absolute-pressure values correct.In an ideal local model, write . If and remain fixed during an effort, thenwhen mouth and pleural pressure changes coincide under the test's assumptions. A ratio near one supports transmission gain near one, but leaves unidentified. Even perfect change agreement cannot determine that constant baseline offset.
The consequence matters when calculating transpulmonary pressure . For an illustrative alveolar pressure of 15 and true local pleural pressure of , the difference is eight. Substituting an esophageal reading of nine with an uncorrected offset of two gives six instead. The arithmetic identifies an error mechanism; it does not supply a clinical correction or assume that all pleural regions have pressure seven.
Pleural-pressure calibration and compensating component errors arise when an internal-pressure estimate partitions total respiratory stiffness into lung and chest-wall contributions. A gain or offset error can shift those inferred components in opposite directions while their sum still agrees perfectly with the measured total.
An original local model treats the volume increment as known. Let true pleural-pressure change be and the measured surrogate beHere is gain and is an offset change between the two volume states. With wall elastance and known total elastance , inference givesThus the component errors are equal and opposite. A constant absolute pressure offset cancels when taking increments, but a changing offset or gain error remains.
For stipulated L, total pressure increment 5 cm water and true pleural increment 2 cm water, the true component elastances are and . A surrogate gain with instead gives 6 and 4. Checking only their sum cannot identify the measurement calibration error. The GUM requires uncertainty and correlations from shared measurement calibration inputs to remain in the measurement model.[ref]
The esophageal balloon occlusion test supplies an experimental constraint. In the 1982 primary study, Ahmet Baydur and colleagues measured changes in esophageal and mouth pressure during inspiratory efforts against an occluded airway in ten subjects and several postures. Pressure-change ratios were near unity in sitting and lateral positions; some supine discrepancies could be reduced by repositioning the balloon.[ref]
A Simple Method for Assessing the Validity of the Esophageal Balloon Technique by Ahmet Baydur and others. Abstract, ten-subject occlusion and posture comparison
The underlying mechanical argument is that an ideal occlusion effort with negligible volume change produces negligible change in elastic transpulmonary pressure. Mouth and pleural pressure changes should then track each other. Comparing the recorded changes tests pressure transmission under those conditions. It does not by itself identify an absolute pressure zero or establish that one position represents every regional pleural pressure during all breathing states. The measurement, its validation maneuver and the inferred elastic partition therefore remain distinct pieces of evidence.