Clinical Experts Group technical article, first publication 14 July 2017, updated 30 September 2020. Opening definition and Calculations distinguish total, external and intrinsic PEEP. Measuring AutoPEEP contrasts a whole-breath least-squares estimate with end-expiratory occlusion and notes potential underestimation by the fitted estimate in severe dynamic hyperinflation. Device-operation instructions are not adopted as individual treatment recommendations.