Fortin barometer 2026-10-31
An adjustable-cistern barometer associated with Jean Nicolas Fortin. Restoring the cistern surface to a fixed index makes the lower mercury level reproducible before the upper level is read.[ref]
A measured mercury column represents pressure through its actual mass density and local gravitational acceleration. An equivalent reference column therefore obeys . If the indicated scale length expands as and mercury volume as , the first-order thermal factor multiplying the scale reading is . These factors describe a pressure-unit reduction, not a physical change in local gravity.[ref]
A mercury barometer balances atmospheric pressure against the hydrostatic head between the open cistern surface and the mercury surface in a closed tube. Writing for the pressure above the column, the ideal relation is
The relevant is the vertical difference between the two mercury levels. It is not simply the length of filled glass, and the top space need not be an absolutely perfect vacuum.
A cistern barometer needs the following corrections or controlled adjustments:
The observational precautions are also described in the Guide to Meteorological Instruments and Methods of Observation by the World Meteorological Organization.[ref] The quantitative corrections are separated in Mercury Barometers and Manometers by W. G. Brombacher and others, particularly the cistern, temperature and gravity discussions. Reduction of a weather-station pressure to sea level is a separate atmospheric comparison, not an intrinsic correction to the instrument.[ref]