Louis A. Shaw, identified in Harvard's institutional timeline as Louis Agassiz Shaw, Jr., coauthored the early tank-respirator reports with Philip Drinker.[ref]
A timeline of Harvard’s history by Harvard University. 1928 entry naming Drinker and Louis Agassiz Shaw, Jr., at the School of Public Health
Their apparatus applied changing pressure around the body while the head remained outside, implementing negative-pressure ventilation rather than requiring the same pressure change at the airway.The second 1929 report, led in the published byline by Shaw, adapted the apparatus for small children and infants. It describes a metal enclosure, rubber neck collar, pump and valve arrangement, with a window for observation. Importantly, the report said the new apparatus had not yet been available for clinical use and cited laboratory tests on curarized cats. Those statements distinguish a reported experimental test from demonstrated clinical outcomes in the intended population.[ref]
An apparatus for the prolonged administration of artificial respiration: II. A design for small children and infants with an appliance for the administration of oxygen and carbon dioxide by Louis A. Shaw and Philip Drinker. Printed pp. 33–35: introductory qualification, enclosure and neck seal
A chamber seal is part of the mechanism. An original steady leakage model lets be suction magnitude relative to room pressure, the chamber-leak conductance and the pump's extraction volumetric flow rate over the modeled operating range. If leakage is linear,at steady state. With hypothetical L/s and , suction is 10 cm HO. Doubling the leak conductance at the same extraction rate reduces it to 5 cm HO. These are illustrative values, not apparatus specifications from the report.
This chamber-leak sensitivity in negative-pressure ventilation separates chamber pumping from pulmonary ventilation. Pumped air through a leaking enclosure is not automatically air passing through the subject's airway. In a further ideal passive mechanical equilibrium model with total respiratory-system compliance 0.01 L/(cm HO), those two driving pressures would produce added volumes 0.10 and 0.05 L. Measuring pump operation alone does not establish delivered lung volume.
The historical research therefore joined physiological reasoning with enclosure geometry, sealing and observation. A successful mechanical cycle tests that the apparatus changes its boundary conditions; independent physiological measurements are needed to establish what breathing assistance those changes provide.
Positive-pressure ventilation 2026-11-05
Positive-pressure ventilation assists inhalation by raising pressure at the airway opening, driving gas into the lungs when that pressure exceeds alveolar pressure. It differs mechanically from spontaneous expansion of the thorax or negative-pressure ventilation, which lowers pressure around the lungs to produce the relevant distending difference. A ventilator's mask or tube supplies the applied airway connection.[ref]
The direction of flow still depends on a pressure difference, not on whether either pressure is above atmospheric. For a simple resistive path,with positive directed inward and the airway resistance. In an original example, pressures 10 and 8 cm HO and cm HO·s/L give inward flow 0.5 L/s. Pressures 8 and 10 cm HO give outward flow 0.5 L/s. Both pairs are positive relative to atmosphere.
This explains why positive end-expiratory pressure can coexist with exhalation. The applied pressure can remain above atmospheric while elastic recoil produces a still higher alveolar pressure and drives gas outward. At equal opening and alveolar pressures, the same ideal resistive model gives zero flow.
The resulting tidal volume also depends on respiratory-system compliance, airway resistance, duration and any active respiratory muscle effort. The no-flow interval used to interpret plateau airway pressure removes one resistive contribution only under the required measurement conditions.[ref] Positive pressure is therefore a mode of applying mechanical assistance, not a guarantee of a particular inward flow, volume or gas-exchange result.